Please use this identifier to cite or link to this item: https://doi.org/10.1115/1.1493803
Title: Multi-axial spine biomechanical testing system with speckle displacement instrumentation
Authors: Chung, S.M. 
Teoh, S.H. 
Tsai, K.T.
Sin, K.K.
Keywords: Biomechanical Testing
Flexion
Multi-Axial
Spine
Issue Date: Aug-2002
Citation: Chung, S.M., Teoh, S.H., Tsai, K.T., Sin, K.K. (2002-08). Multi-axial spine biomechanical testing system with speckle displacement instrumentation. Journal of Biomechanical Engineering 124 (4) : 471-477. ScholarBank@NUS Repository. https://doi.org/10.1115/1.1493803
Abstract: This paper reports on the design and development of a multi-axis (up to 6 axes) mechanical tester for spinal research and testing. The developed spine tester allowed true motion to be simulated on a specimen in pure or combined modes. To demonstrate the capability of the new tester, flexural stiffness properties of sheep lumbar motion segments were evaluated together with a non-contact speckle displacement measurement system. The flexural stiffness of the specimens was measured and compared under constrained and non-constrained testing conditions; with relieving of shear forces (non-constrained), it was found that the specimen behaved in a 'stiffer' manner.
Source Title: Journal of Biomechanical Engineering
URI: http://scholarbank.nus.edu.sg/handle/10635/51466
ISSN: 01480731
DOI: 10.1115/1.1493803
Appears in Collections:Staff Publications

Show full item record
Files in This Item:
There are no files associated with this item.

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.