Please use this identifier to cite or link to this item: https://doi.org/10.1007/s11517-019-01972-5
Title: Design of an online-tuned model based compound controller for a fully automated artificial pancreas
Authors: Bhattacharjee A.
Easwaran A.
Leow M.K.-S. 
Cho N.
Keywords: Artificial pancreas
Enhanced internal model control
Internal model control
Type 1 diabetes mellitus
Volterra model
Issue Date: 2019
Publisher: Springer Verlag
Citation: Bhattacharjee A., Easwaran A., Leow M.K.-S., Cho N. (2019). Design of an online-tuned model based compound controller for a fully automated artificial pancreas. Medical and Biological Engineering and Computing. ScholarBank@NUS Repository. https://doi.org/10.1007/s11517-019-01972-5
Abstract: This paper deals with the development of a control algorithm that can predict optimal insulin doses without patients? intervention in fully automated artificial pancreas system. An online-tuned model based compound controller comprising an online-tuned internal model control (IMC) algorithm and an enhanced IMC (eIMC) algorithm along with a meal detection module is proposed. Volterra models, used to develop IMC and eIMC algorithms, are developed online using recursive least squares (RLS) filter. The time domain kernels, computed online using RLS filter, are converted into frequency domain to obtain Volterra transfer function (VTF). VTFs are used to develop both IMC and eIMC algorithms. The compound controller is designed in such a way that eIMC predicts insulin doses when the glucose rate increase detector of meal detection module is positive, otherwise conventional IMC takes the control action. Experimental results show that the compound controller performs robustly in the presence of higher and irregular amounts of meal disturbances at random times, very high actuator and sensor noises and also with the variation in insulin sensitivity. The combination of compound control strategy and meal detection module compensates the shortcomings of both slow subcutaneous insulin action that causes postprandial hyperglycemia, and delayed peak of action that causes hypoglycaemia. ? 2019, International Federation for Medical and Biological Engineering.
Source Title: Medical and Biological Engineering and Computing
URI: https://scholarbank.nus.edu.sg/handle/10635/177464
ISSN: 01400118
DOI: 10.1007/s11517-019-01972-5
Appears in Collections:Staff Publications

Show full item record
Files in This Item:
There are no files associated with this item.

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.