dc.titleBOD measurement in the presence of heavy metal ions using a thermally-killed-Bacillus subtilis biosensor
dc.contributor.authorQian, Z.
dc.contributor.authorTan, T.C.
dc.date.accessioned2014-10-09T09:51:47Z
dc.date.available2014-10-09T09:51:47Z
dc.date.issued1999-09
dc.identifier.citationQian, Z., Tan, T.C. (1999-09). BOD measurement in the presence of heavy metal ions using a thermally-killed-Bacillus subtilis biosensor. Water Research 33 (13) : 2923-2928. ScholarBank@NUS Repository. https://doi.org/10.1016/S0043-1354(99)00011-1
dc.identifier.issn00431354
dc.identifier.urihttp://scholarbank.nus.edu.sg/handle/10635/91879
dc.description.abstractThe measurement of the biochemical oxygen demand (BOD) of an aqueous solution using a thermally-killed Bacillus subtilis BOD biosensor was unaffected by the presence of 5mM of Al3+, Cd2+, Co2+, Cr3+, Fe2+, Fe3+, Ni2+ and Pb2+ in the sample. In the presence of 5mM of Zn2+, Mn2+ and Sn2+, the sensor over-estimated the BOD5 of the test samples by 12-68% compared with 13% under-estimation at the same concentration of Cu2+. The sensor completely and irreversibly lost its BOD sensing ability in the presence of 0.5mM of Ag+ or Hg2+. Similar effect was observed with a living B. subtilis BOD biosensor. The interference of copper, manganese, stannous and zinc ions was effectively eliminated by dosing the substrate with 6.2mM of ethylene diamine tetra-acetate (EDTA) or 5.8mM of sodium diethyl dithiocarbamate (DDTC). The total concentration of mercury of the solution at which the sensor completely and irreversibly lost its BOD sensing characteristics was raised from 0.5 to 4.7mM when the solution was dosed with 31.8mM of NH4Cl; to 6mM with 6.2mM EDTA and to 3mM with 5.8mM DDTC. The total concentration of silver which would incapacitate the sensor was correspondingly raised to 9.7, 5 and 5.5mM for the respective dosages and complexing agents. The sensor showed no response towards NH4Cl up to about 0.5M and up to about 50mM of EDTA and DDTC. Copyright (C) 1999.
dc.description.urihttp://libproxy1.nus.edu.sg/login?url=http://dx.doi.org/10.1016/S0043-1354(99)00011-1
dc.sourceScopus
dc.subjectBacillus subtilis
dc.subjectBiosensor
dc.subjectBOD
dc.subjectHeavy metal ions
dc.subjectThermally-killed-cells
dc.typeArticle
dc.contributor.departmentCHEMICAL & ENVIRONMENTAL ENGINEERING
dc.description.doi10.1016/S0043-1354(99)00011-1
dc.description.sourcetitleWater Research
dc.description.volume33
dc.description.issue13
dc.description.page2923-2928
dc.description.codenWATRA
dc.identifier.isiut000082510600003
Appears in Collections:Staff Publications

Show simple item record
Files in This Item:
There are no files associated with this item.

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.