dc.titleMonitoring BOD in the presence of heavy metal ions using a poly(4- vinylpyridine)-coated microbial sensor
dc.contributor.authorLi, F.
dc.contributor.authorTan, T.C.
dc.date.accessioned2014-10-09T08:19:52Z
dc.date.available2014-10-09T08:19:52Z
dc.date.issued1994
dc.identifier.citationLi, F., Tan, T.C. (1994). Monitoring BOD in the presence of heavy metal ions using a poly(4- vinylpyridine)-coated microbial sensor. Biosensors and Bioelectronics 9 (6) : 445-455. ScholarBank@NUS Repository. https://doi.org/10.1016/0956-5663(94)90033-7
dc.identifier.issn09565663
dc.identifier.urihttp://scholarbank.nus.edu.sg/handle/10635/91586
dc.description.abstractInterference by heavy metal ions such as Cr3+, Sn2+, Al3+, Co2+, Zn2+, Ni2+, Mn2+, Cu2+ and Ag+, on the response of the Bacillus subtilis/Bacillus licheniformis 7B BOD sensor was effectively eliminated by covering the sensor with a poly(4-vinylpyridine)-coated polycarbonate membrane. The response characteristics of the sensor, with respect to variations in pH, temperature, dissolved alkaline salts, pure organic solutions and wastewater, remained substantially unaffected. The results suggest that a 1 μm thick poly(4-vinylpyridine) film adsorbs heavy metal ions within its structure but at the same time poses no chemical or mass transfer resistance to the diffusion of oxygen or organic solutes from the substrate to the biofilm. Hence, at such thickness, a poly(4-vinylpyridine)- coating eliminates the interference of heavy metal ions with no loss of sensitivity or spectrum of measurable substrates. The mechanism of the elimination is discussed.
dc.description.urihttp://libproxy1.nus.edu.sg/login?url=http://dx.doi.org/10.1016/0956-5663(94)90033-7
dc.sourceScopus
dc.typeArticle
dc.contributor.departmentCHEMICAL ENGINEERING
dc.description.doi10.1016/0956-5663(94)90033-7
dc.description.sourcetitleBiosensors and Bioelectronics
dc.description.volume9
dc.description.issue6
dc.description.page445-455
dc.description.codenBBIOE
dc.identifier.isiutA1994PB13800012
Appears in Collections:Staff Publications

Show simple item record
Files in This Item:
There are no files associated with this item.

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.