Please use this identifier to cite or link to this item: https://scholarbank.nus.edu.sg/handle/10635/62180
DC FieldValue
dc.titleFactors governing the discharge of electrostatic mirror formations in the scanning electron microscope
dc.contributor.authorWong, W.K.
dc.contributor.authorThong, J.T.L.
dc.contributor.authorPhang, J.C.H.
dc.date.accessioned2014-06-17T06:48:15Z
dc.date.available2014-06-17T06:48:15Z
dc.date.issued1998-01
dc.identifier.citationWong, W.K.,Thong, J.T.L.,Phang, J.C.H. (1998-01). Factors governing the discharge of electrostatic mirror formations in the scanning electron microscope. Review of Scientific Instruments 69 (1) : 210-214. ScholarBank@NUS Repository.
dc.identifier.issn00346748
dc.identifier.urihttp://scholarbank.nus.edu.sg/handle/10635/62180
dc.description.abstractTwo primary parameters for an automated electrostatic mirror discharge technique are studied, namely the discharge voltage and the discharge current. It was found that using discharge voltages corresponding to the σ>1 region and near the first crossover point of a sample's total yield curve gave the optimum discharge rate for the technique. Also, using discharge currents larger than the scanning current was shown to improve the rate of discharge. These optimizations ensure the feasibility of this technique as a discharge tool where repeated high and low beam voltage switching in the scanning electron microscope is needed. © 1998 American Institute of Physics.
dc.sourceScopus
dc.typeArticle
dc.contributor.departmentELECTRICAL ENGINEERING
dc.description.sourcetitleReview of Scientific Instruments
dc.description.volume69
dc.description.issue1
dc.description.page210-214
dc.description.codenRSINA
dc.identifier.isiutNOT_IN_WOS
Appears in Collections:Staff Publications

Show simple item record
Files in This Item:
There are no files associated with this item.

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.