Please use this identifier to cite or link to this item: https://doi.org/10.1097/00002480-199310000-00019
Title: In vitro strain measurement of a mechanical heart valve in a pulse simulator
Authors: Teoh, S.H. 
Lee, K.H. 
Nugent, A.H.
Goh, K.S.
Issue Date: 1993
Citation: Teoh, S.H.,Lee, K.H.,Nugent, A.H.,Goh, K.S. (1993). In vitro strain measurement of a mechanical heart valve in a pulse simulator. ASAIO Journal 39 (4) : 929-932. ScholarBank@NUS Repository. https://doi.org/10.1097/00002480-199310000-00019
Abstract: To ascertain the stress magnitude at the stress concentration areas, in vitro strain measurements on a St. Vincent's mechanical heart valve were carried out in a pulse simulator. Results were combined with a finite element (FEM) stress analysis of the titanium valve housing. The maximum stress at the site of stress concentration of the titanium valve housing was 51 MPa. This is well below the fatigue endurance limit of titanium. The imposed stress on the occluder by the upper strut was less than 2 MPa. This is below the lower stress limit of Delrin and may explain why no fracture of the Delrin disk occluder has been reported. The combined use of microstrain analysis and FEM proved to be essential in the determination of dynamic stresses during the opening and closing of the mechanical heart valve.
Source Title: ASAIO Journal
URI: http://scholarbank.nus.edu.sg/handle/10635/58388
ISSN: 10582916
DOI: 10.1097/00002480-199310000-00019
Appears in Collections:Staff Publications

Show full item record
Files in This Item:
There are no files associated with this item.

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.