Please use this identifier to cite or link to this item: https://doi.org/10.1109/TCST.2007.902950
DC FieldValue
dc.titleDesign, modeling, and control of piezoelectric actuators for intracytoplasmic sperm injection
dc.contributor.authorPutra, A.S.
dc.contributor.authorHuang, S.
dc.contributor.authorTan, K.K.
dc.contributor.authorPanda, S.K.
dc.contributor.authorLee, T.H.
dc.date.accessioned2014-06-17T02:44:50Z
dc.date.available2014-06-17T02:44:50Z
dc.date.issued2007
dc.identifier.citationPutra, A.S., Huang, S., Tan, K.K., Panda, S.K., Lee, T.H. (2007). Design, modeling, and control of piezoelectric actuators for intracytoplasmic sperm injection. IEEE Transactions on Control Systems Technology 15 (5 SPEC. ISS.) : 879-890. ScholarBank@NUS Repository. https://doi.org/10.1109/TCST.2007.902950
dc.identifier.issn10636536
dc.identifier.urihttp://scholarbank.nus.edu.sg/handle/10635/55592
dc.description.abstractThe modeling and control of two piezoelectric actuators (PAs), with an objective application towards intracytoplasmic sperm injection (ICSI), is addressed in this paper. The PAs are used to realizing the fast and precise movements of an injector so as to penetrate an oocyte (egg cell) with precision and without causing a large deformation. Two specific PA designs are considered for this application: linear reciprocating and partially rotating. General models are presented for the purpose of control system design for both designs. To address the nonlinearities and uncertainties of the actuators, as represented in the model, an adaptive controller is proposed to yield robust control performance. Application results are presented and discussed in this paper. Successful penetration of the zona pellucida (cell membrane) of the oocyte to the ICSI problem has been achieved with both designs and the merits and demerits of both configuration are duly discussed. © 2007 IEEE.
dc.description.urihttp://libproxy1.nus.edu.sg/login?url=http://dx.doi.org/10.1109/TCST.2007.902950
dc.sourceScopus
dc.subjectAdaptive control
dc.subjectIntracytoplasmic sperm injection (ICSI)
dc.subjectLinear reciprocating
dc.subjectPartially rotating
dc.subjectPiezoelectric actuator (PA)
dc.typeArticle
dc.contributor.departmentELECTRICAL & COMPUTER ENGINEERING
dc.description.doi10.1109/TCST.2007.902950
dc.description.sourcetitleIEEE Transactions on Control Systems Technology
dc.description.volume15
dc.description.issue5 SPEC. ISS.
dc.description.page879-890
dc.description.codenIETTE
dc.identifier.isiut000249161300007