Please use this identifier to cite or link to this item: https://doi.org/10.1109/ACCESS.2019.2909834
DC FieldValue
dc.titleAn Irregular-Shaped Inward-Outward Ring-Pair Magnet Array with a Monotonic Field Gradient for 2D Head Imaging in Low-Field Portable MRI
dc.contributor.authorRen, Z.H.
dc.contributor.authorGong, J.
dc.contributor.authorHuang, S.Y.
dc.date.accessioned2021-12-29T05:46:53Z
dc.date.available2021-12-29T05:46:53Z
dc.date.issued2019
dc.identifier.citationRen, Z.H., Gong, J., Huang, S.Y. (2019). An Irregular-Shaped Inward-Outward Ring-Pair Magnet Array with a Monotonic Field Gradient for 2D Head Imaging in Low-Field Portable MRI. IEEE Access 7 : 48715-48724. ScholarBank@NUS Repository. https://doi.org/10.1109/ACCESS.2019.2909834
dc.identifier.issn2169-3536
dc.identifier.urihttps://scholarbank.nus.edu.sg/handle/10635/212497
dc.description.abstractWe present a design and the optimization of an irregular-shaped inward-outward (IO) ring-pair magnet array that generates a 1D monotonic field pattern for 2D head imaging in a low-field portable magnetic resonance imaging (MRI) system. The magnet rings are discretized into fan-shaped ring segments with varying outer radii for the design and optimization. Besides, the inner radii of ring-pairs are tapered from outside in to provide the controlled field inhomogeneity. Genetic algorithm (GA) was used, and a current model for a fan-shaped ring segment was derived to have a fast forward calculation in the optimization. A monotonic field pattern is successfully obtained along the x-direction in a cylindrical field of view (FoV), with a relatively strong magnetic field (132.98 mT) and the homogeneity of 151840 ppm. The proposed array was further evaluated by applying its field as a spatial encoding magnetic field (SEM) for imaging by using simulation. Due to the field monotonicity, the reconstructed image by applying the fields of the proposed array shows clearer features (a higher structural similarity index) with a reduced error rate compared to that using a sparse dipolar Halbach array. The proposed magnet array is a promising alternative to supply SEM for imaging in a permanent-magnet-based low-field portable MRI system. © 2013 IEEE.
dc.publisherInstitute of Electrical and Electronics Engineers Inc.
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 International
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/4.0/
dc.sourceScopus OA2019
dc.subjectCurrent density model
dc.subjectgenetic algorithm
dc.subjecthomogeneous magnetic field
dc.subjectlinear gradient field
dc.subjectlow-field MRI
dc.subjectmagnet array
dc.subjectpermanent magnet array
dc.subjectportable MRI
dc.subjectspatial encoding field
dc.typeArticle
dc.contributor.departmentSURGERY
dc.description.doi10.1109/ACCESS.2019.2909834
dc.description.sourcetitleIEEE Access
dc.description.volume7
dc.description.page48715-48724
dc.published.statePublished
Appears in Collections:Staff Publications
Elements

Show simple item record
Files in This Item:
File Description SizeFormatAccess SettingsVersion 
10_1109_ACCESS_2019_2909834.pdf14.68 MBAdobe PDF

OPEN

NoneView/Download

Google ScholarTM

Check

Altmetric


This item is licensed under a Creative Commons License Creative Commons