dc.titleA method to assess image quality for Low-dose PET: Analysis of SNR, CNR, bias and image noise
dc.contributor.authorYan, J
dc.contributor.authorSchaefferkoette, J
dc.contributor.authorConti, M
dc.contributor.authorTownsend, D
dc.date.accessioned2020-09-14T08:12:11Z
dc.date.available2020-09-14T08:12:11Z
dc.date.issued2016
dc.identifier.citationYan, J, Schaefferkoette, J, Conti, M, Townsend, D (2016). A method to assess image quality for Low-dose PET: Analysis of SNR, CNR, bias and image noise. Cancer Imaging 16 (1) : 26. ScholarBank@NUS Repository. https://doi.org/10.1186/s40644-016-0086-0
dc.identifier.issn1470-7330
dc.identifier.urihttps://scholarbank.nus.edu.sg/handle/10635/176121
dc.description.abstractBackground: Lowering injected dose will have an effect on PET image quality. In this article, we aim to investigate this effect in terms of signal-to-noise ratio (SNR) in the liver, contrast-to-noise ratio (CNR) in the lesion, bias and ensemble image noise. Methods: We present here our method and preliminary results using tuberculosis (TB) cases. Sixteen patients who underwent 18F-FDG PET/MR scans covering the whole lung and portion of the liver were selected for the study. Reduced doses were simulated by randomly discarding events in the PET list mode data stream, and ten realizations at each simulated dose were generated and reconstructed. The volumes of interest (VOI) were delineated on the image reconstructed from the original full statistics data for each patient. Four thresholds (20, 40, 60 and 80 % of SUVmax) were used to quantify the effect of the threshold on CNR at the different count level. Image metrics were calculated for each VOI. This experiment allowed us to quantify the loss of SNR and CNR as a function of the counts in the scan, in turn related to dose injected. Reproducibility of mean and maximum standardized uptake value (SUVmean and SUVmax) measurement in the lesions was studied as standard deviation across 10 realizations. Results: At 5 × 106 counts in the scan, the average SNR in the liver in the observed samples is about 3, and the CNR is reduced to 60 % of the full statistics value. The CNR in the lesion and SNR in the liver decreased with reducing count data. The variation of CNR across the four thresholds does not significantly change until the count level of 5 × 106. After correcting the factor related to subject's weight, the square of the SNR in the liver was found to have a very good linear relationship with detected counts. Some quantitative bias appears with count reduction. At the count level of 5 × 106, bias and noise in terms of SUVmean and SUVmax are up to 10 and 20 %, respectively. To keep both bias and noise less than 10 %, 5 × 106 counts and 20 × 106 counts were required for SUVmean and SUVmax, respectively. Conclusions: Initial results with the given data of 16 patients diagnosed as TB demonstrated that 5 × 106 counts in the scan could be sufficient to yield good images in terms of SNR, CNR, bias and noise. In the future, more work needs to be done to validate the proposed method with a larger population and lung cancer patient data. © 2016 The Author(s).
dc.sourceUnpaywall 20200831
dc.subjectfluorodeoxyglucose f 18
dc.subjectfluorodeoxyglucose f 18
dc.subjectradiopharmaceutical agent
dc.subjectadult
dc.subjectaged
dc.subjectArticle
dc.subjectcancer screening
dc.subjectclinical article
dc.subjectcomputer assisted emission tomography
dc.subjectcontrast to noise ratio
dc.subjectfemale
dc.subjecthuman
dc.subjectimage analysis
dc.subjectimage display
dc.subjectimage processing
dc.subjectimage quality
dc.subjectliver
dc.subjectlung cancer
dc.subjectmale
dc.subjectquantitative analysis
dc.subjectradiological parameters
dc.subjectreproducibility
dc.subjectsignal noise ratio
dc.subjecttuberculosis
dc.subjectpositron emission tomography
dc.subjectprocedures
dc.subjectradiation response
dc.subjectsignal noise ratio
dc.subjectDose-Response Relationship, Radiation
dc.subjectFluorodeoxyglucose F18
dc.subjectHumans
dc.subjectPositron-Emission Tomography
dc.subjectRadiopharmaceuticals
dc.subjectReproducibility of Results
dc.subjectSignal-To-Noise Ratio
dc.typeArticle
dc.contributor.departmentDIAGNOSTIC RADIOLOGY
dc.description.doi10.1186/s40644-016-0086-0
dc.description.sourcetitleCancer Imaging
dc.description.volume16
dc.description.issue1
dc.description.page26
dc.published.statePublished
Appears in Collections:Elements
Staff Publications

Show simple item record
Files in This Item:
File Description SizeFormatAccess SettingsVersion 
10_1186_s40644-016-0086-0.pdf2.3 MBAdobe PDF

OPEN

NoneView/Download

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.