dc.titleCorrection of energy-dependent systematic errors in dual-energy X-ray CT using a basis material coefficients transformation method
dc.contributor.authorGoh, K.L.
dc.contributor.authorLiew, S.C.
dc.contributor.authorHasegawa, B.H.
dc.date.accessioned2014-11-28T07:57:36Z
dc.date.available2014-11-28T07:57:36Z
dc.date.issued1996
dc.identifier.citationGoh, K.L.,Liew, S.C.,Hasegawa, B.H. (1996). Correction of energy-dependent systematic errors in dual-energy X-ray CT using a basis material coefficients transformation method. IEEE Nuclear Science Symposium & Medical Imaging Conference 2 : 1453-1457. ScholarBank@NUS Repository.
dc.identifier.urihttp://scholarbank.nus.edu.sg/handle/10635/112858
dc.description.abstractComputer simulation results from our previous studies showed that energy dependent systematic errors exist in the values of attenuation coefficient synthesized using the Basis material decomposition technique with acrylic and aluminum as the basis materials, especially when a high atomic number element (e.g., iodine from radiographic contrast media) was present in the body. The errors were reduced when a basis set was chosen from materials mimicking those found in the phantom. In the present study, we employed a basis material coefficients transformation method to correct for the energy-dependent systematic errors. In this method, the basis material coefficients were first reconstructed using the conventional basis materials (acrylic and aluminum) as the calibration basis set. The coefficients were then numerically transformed to those for a more desirable set of basis materials. The transformation was done at the effective energies of the low and high energy windows of the x-ray spectrum. With this correction method using acrylic and an iodine-water mixture as our desired basis set, computer simulation results showed that accuracy of better than 2% could be achieved even when iodine was present in the body at a concentration as high as 10% by mass.
dc.sourceScopus
dc.typeConference Paper
dc.contributor.departmentCTR FOR REM IMAGING,SENSING & PROCESSING
dc.description.sourcetitleIEEE Nuclear Science Symposium & Medical Imaging Conference
dc.description.volume2
dc.description.page1453-1457
dc.description.coden85OQA