dc.titleCalibration sampling paradox in near infrared spectroscopy: A case study of multi-component powder blend
dc.contributor.authorKarande, A.D.
dc.contributor.authorLiew, C.V.
dc.contributor.authorHeng, P.W.S.
dc.date.accessioned2014-10-29T01:49:30Z
dc.date.available2014-10-29T01:49:30Z
dc.date.issued2010-08
dc.identifier.citationKarande, A.D., Liew, C.V., Heng, P.W.S. (2010-08). Calibration sampling paradox in near infrared spectroscopy: A case study of multi-component powder blend. International Journal of Pharmaceutics 395 (1-2) : 91-97. ScholarBank@NUS Repository. https://doi.org/10.1016/j.ijpharm.2010.05.011
dc.identifier.issn03785173
dc.identifier.urihttp://scholarbank.nus.edu.sg/handle/10635/105711
dc.description.abstractThe objective of this study was to illustrate the sampling paradox resulting from the different strategies of spectral acquisition while preparing and implementing the calibration models for prediction of blend components in multi-component cohesive blends. A D-optimal mixture design was used to create 24 blending runs of the formulation consisting of chlorpheniramine maleate, lactose, microcrystalline cellulose and magnesium stearate. Three strategies: (a) laboratory mixing and static spectral acquisition, (b) IBC mixing and static spectral acquisition and (c) IBC mixing and dynamic spectral acquisition were investigated for obtaining the most relevant and representative calibration samples. An optical head comprising a sapphire window mounted on the lid of the IBC was used for static and dynamic NIR spectral acquisition of the powder blends. For laboratory mixed samples, powders were blended for fixed period of 30. min and later on scanned for NIR spectra. For IBC mixed blends, the spectral acquisition was carried out in-line for 2. min and stopped for static spectral acquisition. The same cycle was repeated for the next 28. min. Partial least square (PLS) calibration models for each component were built and ranked according to their calibration statistics. Optimal calibration models were selected from each strategy for each component and used for in-line prediction of blend components of three independent test runs. Although excellent statistics were obtained for the PLS models from the three strategies, significant discrepancies were observed during prediction of the independent blends in real time. Models built using IBC mixed blends and dynamic spectral acquisition resulted in the most accurate predictions for all the blend components, whereas models prepared using static spectral acquisition (laboratory mixed and IBC) showed erroneous prediction results. The prediction performance differences between the models obtained using the different strategies could be explained in the context of relevancy and representative sample collection at the initial stage of calibration model building. © 2010 Elsevier B.V.
dc.description.urihttp://libproxy1.nus.edu.sg/login?url=http://dx.doi.org/10.1016/j.ijpharm.2010.05.011
dc.sourceScopus
dc.subjectBlending
dc.subjectDynamic spectral acquisition
dc.subjectMultivariate calibration models
dc.subjectNIR
dc.subjectSampling
dc.subjectStatic spectral acquisition
dc.typeArticle
dc.contributor.departmentPHARMACY
dc.description.doi10.1016/j.ijpharm.2010.05.011
dc.description.sourcetitleInternational Journal of Pharmaceutics
dc.description.volume395
dc.description.issue1-2
dc.description.page91-97
dc.description.codenIJPHD
dc.identifier.isiut000280212500013
Appears in Collections:Staff Publications

Show simple item record
Files in This Item:
There are no files associated with this item.

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.