Please use this identifier to cite or link to this item:
https://doi.org/10.1117/12.696623
Title: | An electromechanical filter for bioparticles trapping | Authors: | Iliescu, C. Xu, G. Tay, F.E.H. Ong, P.L. |
Keywords: | Dielectrophoresis Filter Microfabrication Particle trapping |
Issue Date: | 2007 | Citation: | Iliescu, C., Xu, G., Tay, F.E.H., Ong, P.L. (2007). An electromechanical filter for bioparticles trapping. Proceedings of SPIE - The International Society for Optical Engineering 6416 : -. ScholarBank@NUS Repository. https://doi.org/10.1117/12.696623 | Abstract: | This work presents the microfabrication procedures and filtering application of a novel 3-D dielectrophoretic chip that possesses a structure similar to a classical capacitor. It is made up by bonding two stainless steel meshes on the opposite sides of a glass frame which is filled with round silica beads. Double filtration actions that are derived from both mechanical and dielectrophoretic means have been tested with yeast cells and a maximum trapping efficiency of approximately 75% has been achieved with initial concentration of 5×106 cells/ml. This was done at an applied voltage of 200 V and a flow rate of 0.1ml/min. | Source Title: | Proceedings of SPIE - The International Society for Optical Engineering | URI: | http://scholarbank.nus.edu.sg/handle/10635/85876 | ISBN: | 0819465240 | ISSN: | 0277786X | DOI: | 10.1117/12.696623 |
Appears in Collections: | Staff Publications |
Show full item record
Files in This Item:
There are no files associated with this item.
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.