Please use this identifier to cite or link to this item: https://scholarbank.nus.edu.sg/handle/10635/62155
Title: Error voltage components in quantitative voltage contrast measurement systems
Authors: Chan, D.S.H. 
Low, T.S. 
Chim, W.K. 
Phang, J.C.H. 
Issue Date: Jun-1991
Citation: Chan, D.S.H.,Low, T.S.,Chim, W.K.,Phang, J.C.H. (1991-06). Error voltage components in quantitative voltage contrast measurement systems. Scanning Microscopy 5 (2) : 345-355. ScholarBank@NUS Repository.
Abstract: This paper presents the results of computer simulation studies into the respective contributions of the potential barrier, the off-normal incidence injection of secondary electrons (SEs) into the retarding field and analyser geometry on Types I and II local field error voltages for a practical 20 mm wide planar retarding field energy analyser. Results show that the error voltage component due to the off-normal incidence injection effect of SEs into the retarding field dominates the Type I local field error. For type II LFE, the error voltage component due to analyser geometry effect is the higher contributing factor. The presence of a neighbouring electrode voltage tends to draw SEs away from the central axis of the energy analyser, thus causing the electron trajectories to be more sensitive to the influence of the analyser geometry.
Source Title: Scanning Microscopy
URI: http://scholarbank.nus.edu.sg/handle/10635/62155
ISSN: 08917035
Appears in Collections:Staff Publications

Show full item record
Files in This Item:
There are no files associated with this item.

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.