Title: | Calibrating an interferometric laser frequency stabilization to megahertz precision | Authors: | Brachmann, J.F.S. Kinder, T. Dieckmann, K. |
Issue Date: | 1-Aug-2012 | Citation: | Brachmann, J.F.S., Kinder, T., Dieckmann, K. (2012-08-01). Calibrating an interferometric laser frequency stabilization to megahertz precision. Applied Optics 51 (22) : 5517-5521. ScholarBank@NUS Repository. https://doi.org/10.1364/AO.51.005517 | Abstract: | We report on a calibration procedure that enhances the precision of an interferometer based frequency stabilization by several orders of magnitude. For this purpose, the frequency deviations of the stabilization are measured precisely by means of a frequency comb. This allows us to implement several calibration steps that compensate different systematic errors. The resulting frequency deviation is shown to be less than 5.7 MHz (rms 1.6 MHz) in the whole wavelength interval 750-795 nm. Wide tuning of a stabilized laser at this exceptional precision is demonstrated. © 2012 Optical Society of America. | Source Title: | Applied Optics | URI: | http://scholarbank.nus.edu.sg/handle/10635/95916 | ISSN: | 1559128X | DOI: | 10.1364/AO.51.005517 |