Please use this identifier to cite or link to this item: https://scholarbank.nus.edu.sg/handle/10635/62025
Title: Detection of 180° phase discrepancies in multi-port waveguide transmission measurements
Authors: Yeo, S P 
Cheng, M. 
Bannister, D J
Ide, J P
Issue Date: 1994
Citation: Yeo, S P,Cheng, M.,Bannister, D J,Ide, J P (1994). Detection of 180° phase discrepancies in multi-port waveguide transmission measurements. Conference Record - IEEE Instrumentation and Measurement Technology Conference 2 : 874-877. ScholarBank@NUS Repository.
Abstract: It has recently been observed that improper choice of relative adaptor orientations may cause 180° phase discrepancies to appear in the measurement data obtained by automatic network analyzers for the transmission parameters of multi-port rectangular-waveguide components. The present paper describes the use of two different test procedures - based either on the principal subdeterminants of the dissipation matrix or on the maximum singular value of the scattering matrix - to detect the presence of these discrepancies.
Source Title: Conference Record - IEEE Instrumentation and Measurement Technology Conference
URI: http://scholarbank.nus.edu.sg/handle/10635/62025
Appears in Collections:Staff Publications

Show full item record
Files in This Item:
There are no files associated with this item.

Page view(s)

85
checked on Mar 30, 2023

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.