Please use this identifier to cite or link to this item: https://doi.org/10.1109/TMI.2010.2045767
DC FieldValue
dc.titleAccurate measurement of bone mineral density using clinical CT imaging with single energy beam spectral intensity correction
dc.contributor.authorZhang, J.
dc.contributor.authorYan, C.-H.
dc.contributor.authorChui, C.-K.
dc.contributor.authorOng, S.H.
dc.date.accessioned2014-04-24T07:19:28Z
dc.date.available2014-04-24T07:19:28Z
dc.date.issued2010-07
dc.identifier.citationZhang, J., Yan, C.-H., Chui, C.-K., Ong, S.H. (2010-07). Accurate measurement of bone mineral density using clinical CT imaging with single energy beam spectral intensity correction. IEEE Transactions on Medical Imaging 29 (7) : 1382-1389. ScholarBank@NUS Repository. https://doi.org/10.1109/TMI.2010.2045767
dc.identifier.issn02780062
dc.identifier.urihttp://scholarbank.nus.edu.sg/handle/10635/50858
dc.description.abstractAlthough dual-energy X-ray absorptiometry (DXA) offers an effective measurement of bone mineral density, it only provides a 2-D projected measurement of the bone mineral density. Clinical computed tomography (CT) imaging will have to be employed for measurement of 3-D bone mineral density. The typical dual energy process requires precise measurement of the beam spectral intensity at the 80 kVp and 120 kVp settings. However, this is not used clinically because of the extra radiation dosage and sophisticated hardware setup. We propose an accurate and fast approach to measure bone material properties with single energy scans. Beam hardening artifacts are eliminated by incorporating the polychromatic characteristics of the X-ray beam into the reconstruction process. Bone mineral measurement from single energy CT correction is compared with that of dual energy correction and the commonly used DXA. Experimental results show that single energy correction is compatible with dual energy CT correction in eliminating beam hardening artifacts and producing an accurate measurement of bone mineral density. We can then estimate Young's modulus, yield stress, yield strain and ultimate tensile stress of the bone, which are important data for patient specific therapy planning. © 2006 IEEE.
dc.description.urihttp://libproxy1.nus.edu.sg/login?url=http://dx.doi.org/10.1109/TMI.2010.2045767
dc.sourceScopus
dc.subjectBeam hardening correction
dc.subjectbeam spectrum
dc.subjectbone mineral density
dc.subjectcomputed tomography
dc.subjectpatient specific therapy
dc.typeArticle
dc.contributor.departmentMECHANICAL ENGINEERING
dc.contributor.departmentELECTRICAL & COMPUTER ENGINEERING
dc.description.doi10.1109/TMI.2010.2045767
dc.description.sourcetitleIEEE Transactions on Medical Imaging
dc.description.volume29
dc.description.issue7
dc.description.page1382-1389
dc.description.codenITMID
dc.identifier.isiut000281925700004
Appears in Collections:Staff Publications

Show simple item record
Files in This Item:
There are no files associated with this item.

SCOPUSTM   
Citations

14
checked on Mar 15, 2023

WEB OF SCIENCETM
Citations

11
checked on Mar 15, 2023

Page view(s)

180
checked on Mar 16, 2023

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.