Title: Effect of vessel enhancement filters on the repeatability of measurements obtained from widefield swept-source optical coherence tomography angiography
Authors: Hong, J.
Ke, M.
Tan, B.
Lau, A.
Wong, D.
Yao, X.
Liu, X.
Schmetterer, L. 
Chua, J. 
Issue Date: 2020
Publisher: Nature Research
Citation: Hong, J., Ke, M., Tan, B., Lau, A., Wong, D., Yao, X., Liu, X., Schmetterer, L., Chua, J. (2020). Effect of vessel enhancement filters on the repeatability of measurements obtained from widefield swept-source optical coherence tomography angiography. Scientific Reports 10 (1) : 22179. ScholarBank@NUS Repository. https://doi.org/10.1038/s41598-020-79281-3
Rights: Attribution 4.0 International
Abstract: We assessed the inter-visit repeatability of 15 × 9-mm2 swept-source OCTA (SS-OCTA; PLEX Elite 9000, Carl Zeiss Meditec) metrics in 14 healthy participants. We analysed the perfusion density (PD) of large vessels, superficial capillary plexus (SCP), and deep capillary plexus (DCP) as well as choriocapillaris flow voids in 2 different regions: the macular region and peripheral region. Also, retinal plexus metrics were processed further using different filters (Hessian, Gabor and Bayesian) while choriocapillaris flow voids were calculated with 1 and 1.25 standard deviation (SD) thresholding algorithms. We found excellent repeatability in the perfusion densities of large vessels (ICC > 0.96). Perfusion densities varied with different filters in the macular region (SCP: 24.12–38.57% and DCP: 25.16–38.50%) and peripheral (SCP: 30.52–39.84% and DCP: 34.19–41.60%) regions. The ICCs were lower in the macular region compared to the peripheral region and lower for DCP than for SCP. For choriocapillaris flow voids, the 1.25 SD threshold resulted in fewer flow voids, while a good ICC (ICC > 0.81) was achieved using either threshold settings for flow void features in both regions. Our results suggest good repeatability of widefield SS-OCTA for the measurements of retinal perfusion density and choriocapillaris flow voids, but measurements from different filters should not be interchanged. © 2020, The Author(s).
Source Title: Scientific Reports
URI: https://scholarbank.nus.edu.sg/handle/10635/196313
ISSN: 2045-2322
DOI: 10.1038/s41598-020-79281-3
Rights: Attribution 4.0 International
Appears in Collections:Elements
Staff Publications

Show full item record
Files in This Item:
File Description SizeFormatAccess SettingsVersion 
10_1038_s41598_020_79281_3.pdf8.4 MBAdobe PDF

OPEN

NoneView/Download

Google ScholarTM

Check

Altmetric


This item is licensed under a Creative Commons License Creative Commons