Title: Differential Pulse Cathodic Stripping Voltammetric Determination of Manganese(II) and Manganese(VII) at the 1-(2-Pyridylazo)-2-Naphthol-Modified Carbon Paste Electrode
Authors: Khoo, S.B. 
Soh, M.K.
Cai, Q.
Khan, M.R.
Guo, S.X.
Keywords: 1-(2- Pyridylazo)-2-Naphthol
Manganese
Modified electrodes
Stripping voltammetry
Issue Date: Jan-1997
Source: Khoo, S.B.,Soh, M.K.,Cai, Q.,Khan, M.R.,Guo, S.X. (1997-01). Differential Pulse Cathodic Stripping Voltammetric Determination of Manganese(II) and Manganese(VII) at the 1-(2-Pyridylazo)-2-Naphthol-Modified Carbon Paste Electrode. Electroanalysis 9 (1) : 45-51. ScholarBank@NUS Repository.
Abstract: A carbon paste electrode modified with 1-(2-pyridylazo)-2-naphthol was fabricated and used to preferentially accumulate Mn(II) at open circuit. After medium exchange, the accumulated Mn(II) at the electrode surface was electrochemically oxidized. The hydrated MnO2 thus formed was then cathodically stripped using differential pulse voltammetry. Various factors influencing the accumulation and stripping were studied. An optimized procedure was then developed for the determination of Mn(II). Mn(VII) could also be determined by chemical conversion to Mn(II) using acidified H2O2. Following this, the Mn(VII) concentration could be obtained by subtraction. Under optimum conditions, a detection limit of 6.9 × 10-9 M Mn(II) (0.38 ppb) was found for 200 s accumulation. For 8 determinations, each of Mn(II) concentrations of 1.00 × 10-6 M, 1.00 × 10-7 M and 1.00 × 10-8 M, relative standard deviations of 2.90 %, 4.46 % and 7.20 % were obtained, respectively. Selected metal ions were tested for interferences and it was found that, when determining 1.00 × 10-6 M Mn(II) (150 s accumulation), Co(II), Hg(II) and Fe(II) at 1.00 × 10-5 M interfered. These interferences were readily eliminated by masking with complexing agents such as sodium citrate and sodium diethyldithiocarbamate. The developed method was tested using a certified sample and then applied to the determination of manganese in seawater.
Source Title: Electroanalysis
URI: http://scholarbank.nus.edu.sg/handle/10635/93586
ISSN: 10400397
Appears in Collections:Staff Publications

Show full item record
Files in This Item:
There are no files associated with this item.

Page view(s)

25
checked on Feb 23, 2018

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.