Title: Analysis of endocrine disrupting alkylphenols, chlorophenols and bisphenol-A using hollow fiber-protected liquid-phase microextraction coupled with injection port-derivatization gas chromatography-mass spectrometry
Authors: Basheer, C. 
Lee, H.K. 
Keywords: Alkylphenols
Bisphenol-A
Derivatization
Hollow fiber membrane
Liquid-phase microextraction
Seawater analysis
Issue Date: 19-Nov-2004
Citation: Basheer, C., Lee, H.K. (2004-11-19). Analysis of endocrine disrupting alkylphenols, chlorophenols and bisphenol-A using hollow fiber-protected liquid-phase microextraction coupled with injection port-derivatization gas chromatography-mass spectrometry. Journal of Chromatography A 1057 (1-2) : 163-169. ScholarBank@NUS Repository. https://doi.org/10.1016/j.chroma.2004.09.083
Abstract: Liquid-phase microextraction (LPME) coupled with gas chromatography-mass spectrometry were used to determine alkylphenols (APs), chlorophenols (CPs) and bisphenol-A (BPA) in aqueous samples. APs, CPs and BPA are highly polar compounds and need to be derivatized before analysis by GC-MS. In this work, they were derivatized in the GC injection port with bis(trimethylsilyl) trifluoroacetamide (BSTFA). The analytes were extracted directly from 5 ml of sample solution using 5 μl of organic solvent though a porous polypropylene hollow fiber. The hollow fiber, filled with an immiscible organic solvent (ca. 5 μl), was immersed in the sample solution which was stirred during the 30-min extraction. An aliquot (2 μl) of the extract and 2 μl of BSTFA were then consecutively injected into the GC injection port. Extraction parameters such as extraction time, pH of sample, concentration of salt added, and stirring rate were optimised. The proposed LPME provided a good average enrichment factor of up to 162-fold, reproducibility ranging from 5.9 to 13.9% (n = 4), and good linearity (r 2 = 0.995) for spiked water samples. The limits of detection (LODs) ranged between 0.005 and 0.015 μg l -1 (S/N = 3) using GC-MS with selective ion monitoring and limits of quantification were in the range of 0.012-0.026 μg l -1. A comparative study was performed between LPME, headspace solid-phase microextraction (HS-SPME) and liquid-liquid extraction (LLE). The results obtained suggested that hollow fiber LPME was a rapid, simple and efficient technique for APs, CPs and BPA, and provided a good alternative to SPME and LLE. Finally, the proposed method was applied to monitor Singapore coastal water samples. © 2004 Elsevier B.V. All rights reserved.
Source Title: Journal of Chromatography A
URI: http://scholarbank.nus.edu.sg/handle/10635/93099
ISSN: 00219673
DOI: 10.1016/j.chroma.2004.09.083
Appears in Collections:Staff Publications

Show full item record
Files in This Item:
There are no files associated with this item.

SCOPUSTM   
Citations

157
checked on Sep 19, 2018

WEB OF SCIENCETM
Citations

151
checked on Sep 19, 2018

Page view(s)

28
checked on Aug 17, 2018

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.