Please use this identifier to cite or link to this item: https://doi.org/10.1109/DELTA.2010.43
DC FieldValue
dc.titleAn ECG-on-chip for wearable cardiac monitoring devices
dc.contributor.authorDeepu, C.J.
dc.contributor.authorXu, X.Y.
dc.contributor.authorZou, X.D.
dc.contributor.authorYao, L.B.
dc.contributor.authorLian, Y.
dc.date.accessioned2014-10-07T04:41:32Z
dc.date.available2014-10-07T04:41:32Z
dc.date.issued2010
dc.identifier.citationDeepu, C.J., Xu, X.Y., Zou, X.D., Yao, L.B., Lian, Y. (2010). An ECG-on-chip for wearable cardiac monitoring devices. Proceedings - 5th IEEE International Symposium on Electronic Design, Test and Applications, DELTA 2010 : 225-228. ScholarBank@NUS Repository. https://doi.org/10.1109/DELTA.2010.43
dc.identifier.isbn9780769539782
dc.identifier.urihttp://scholarbank.nus.edu.sg/handle/10635/83466
dc.description.abstractThis paper describes a highly integrated, low power chip solution for ECG signal processing in wearable devices. The chip contains an instrumentation amplifier with programmable gain, a band-pass filter, a 12-bit SAR ADC, a novel QRS detector, 8K on-chip SRAM, and relevant control circuitry and CPU interfaces. The analog front end circuits accurately senses and digitizes the raw ECG signal, which is then filtered to extract the QRS. The sampling frequency used is 256 Hz. ECG samples are buffered locally on an asynchronous FIFO and is read out using a faster clock, as and when it is required by the host CPU via an SPI interface. The chip was designed and implemented in 0.35μm standard CMOS process. The analog core operates at 1V while the digital circuits and SRAM operate at 3.3V. The chip total core area is 5.74 mm2 and consumes 9.6μW. Small size and low power consumption make this design suitable for usage in wearable heart monitoring devices. © 2010 IEEE.
dc.description.urihttp://libproxy1.nus.edu.sg/login?url=http://dx.doi.org/10.1109/DELTA.2010.43
dc.sourceScopus
dc.subjectECG-on-chip
dc.subjectElectrocardiography
dc.subjectLow power design
dc.subjectQRS detection
dc.subjectWearable devices
dc.typeConference Paper
dc.contributor.departmentELECTRICAL & COMPUTER ENGINEERING
dc.description.doi10.1109/DELTA.2010.43
dc.description.sourcetitleProceedings - 5th IEEE International Symposium on Electronic Design, Test and Applications, DELTA 2010
dc.description.page225-228
dc.identifier.isiutNOT_IN_WOS
Appears in Collections:Staff Publications
Elements

Show simple item record
Files in This Item:
File Description SizeFormatAccess SettingsVersion 
2010-an_ecg-on-chip_for_wearable_cardiac-postprint.pdf612.65 kBAdobe PDF

CLOSED (no policy)

Post-print

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.