Title: The impact of off the visual axis retinoscopy on objective central refractive measurement in adult clinical practice: A prospective, randomized clinical study
Authors: Tay, E.
Mengher, L.
Lin, X.-Y. 
Ferguson, V.
Keywords: error
off axis
power
refractive
retinoscopy
vectors
Issue Date: Jul-2011
Citation: Tay, E., Mengher, L., Lin, X.-Y., Ferguson, V. (2011-07). The impact of off the visual axis retinoscopy on objective central refractive measurement in adult clinical practice: A prospective, randomized clinical study. Eye 25 (7) : 888-892. ScholarBank@NUS Repository. https://doi.org/10.1038/eye.2011.79
Abstract: Purpose The objective of this study was to examine the effect of the off axis retinoscopy on objective central refractive measurement in adult clinical practice. Methods In all, 40 subjects underwent undilated retinoscopy in a randomly selected eye both on and off the visual axis by a single masked examiner. Off axis retinoscopy was defined as retinoscopy performed with the testing eye of the examiner aligned with the contralateral (non-test) eye of the subject resulting in an off axis deviation in the nasal horizontal visual field. Retinoscopy was performed in negative cylinder only and spherocylindrical measurements were converted to power vectors for analysis. Paired t-test was used to assess differences in M, J0 and J45 power vectors including differences between mean aided and unaided LogMar acuities. Results In all, 14 subjects were myopic (SE ≤ 0.5 D), 13 subjects were emmetropic (SE between -0.49 and 1.0 D) and 13 subjects were hyperopic (SE ≥ 1.0 D). Mean angle of deviation was 5.58° in the nasal horizontal visual field. J 0 showed a significant negative shift in those with myopia (P≤0.001) and emmetropia (P=0.049) following off axis retinoscopy. No significant differences in M, J0 and J45 were found in the hyperopes. Mean aided LogMar acuities after on and off axis retinoscopy were both significantly better than mean unaided LogMar VA (P≤0.001 in both cases). Conclusion Small degrees of off axis retinoscopy encountered in everyday clinical practice can induce errors in objective central refractive measurement. © 2011 Macmillan Publishers Limited All rights reserved.
Source Title: Eye
URI: http://scholarbank.nus.edu.sg/handle/10635/108574
ISSN: 0950222X
DOI: 10.1038/eye.2011.79
Appears in Collections:Staff Publications

Show full item record
Files in This Item:
There are no files associated with this item.

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.